Miniflash FP Vision Flash Point Analyzer

Flash point analyzer for solid and liquid samples with maximum safety.

Category:
  1. General Details

    The flash point tester MINIFLASH FP Visionis uniquely designed to determine the flash points of liquids and solids. The analyzer uses Grabner flash point detection method, which is triggered by an energy-controlled electric arc and measures instantaneous pressure increase inside a closed chamber. By heating the test chamber from the top, the analyzer prevents condensation of high volatile compounds and guarantees improved test results.


    MINIFLASH FP Vision continuously closed cup design and small sample volume of only 1-2 ml offer maximum safety. No offensive fumes are released in the testing area and little to no sample waste is registered when performing a flash point test. The fast thermoelectric cooling at the end of each flash point test plus the easy handling of the instrument also save expensive labor time.

  2. Technical Info
    Temperature Range

     

     

    FP Vision:

    • 0 to 120°C (32 to 248°F) without cooling

    • Down to -25°C (-13°F) with water cooling

    FPH Vision:

    • 10 to 400°C (50 to 752°F)

    Temperature Stability

     

    FP Vision: +/- 0.05°C (0.09 °F)

    FPH Vision: +/- 0.07 °C (0.13 °F)

    SampleVolume 1 ml (ASTM D6450) or 2 ml (ASTM D7094)
    Fast Sample ThroughputUp to 16 samples/hour, depending on method
    Interfaces

     

    4x USB, 2x LAN, Remote Control Remote Control via COCKPITTM Software for Vision analyzers
    Power Supply100/110/120/230/240 V AC, 50/60 Hz, max. 180W

    (optional car adapter for field use)

    Dimensions(WxHxD) 253 x 368 x 277 mm (10 x 14.5 x 10.9 inch)
    Weight10.2-11.2kg (22.4-lb)
  3. Order Inquiry