LaserNet 200 Oil Particle Counter & Ferrous Content Analyzer

Fast, accurate particle count and shape classification in lubricants, fuels and similar liquids.

  1. General Details

    LaserNet 200 Series apply the world’s best oil particle counter technology.  With a configurable GUI, LaserNet provides fast, accurate and easy to use analysis. Its unique imaging technology eliminates the need of calibration for lifetime.

    Sample preparation is efficient with the LaserNet. Viscosities up to 320 cSt can be processed without dilution. Unlike conventional light blockage particle counters, there are no flow control valves that need adjusting when testing different sample viscosities.

    The advanced design makes it much more than just a particle counter. In addition, LaserNet also calculates free water in ppm, and differentiates contaminants. Several models are available providing particle counts and codes, abnormal wear classification and ferrous wear measurement.


    Oil Particle Counter Features & Benefits

    • ASTM D7596 – Standard Test Method for Automatic Particle Counting and Particle Shape Classification of Oils Using a Direct Imaging Integrated Tester
    • Particle Count for all particles >4 μm to 100 μm
    • Highest saturation limit, up to 5,000,000 particles/ml with a coincidence error <2%
    • Viscosity range ISO15 to ISO320, undiluted
    • Images through dark fluids containing up to 2% soot with automatic laser gain control
    • Error corrections for water and air bubbles
    • Particle counts and codes per ISO 4406, NAS 1638, NAVAIR 01-1A-17, SAE AS 4059, GOST, STM D6786, HAL, and User defined bins.
    • Export formats include TruVu 360 and AMS Machinery Health Manager

    Classification of Wear Particle Shape

    LaserNet Oil Particle Counter directly images wear particle silhouettes. LaserNet counts and classifies wear particles over 20 microns as cutting wear, sliding wear, fatigue wear, nonmetallic or fibers. This allows operators to determine the type of wear debris, wear mode and condition of machinery components.


    Ferrous Content Measurement

    Ferrous wear measurement is a critical requirement for monitoring machine condition. The high sensitivity magnetometer measures and reports ferrous content in ppm/ml. This unit provides ferrous particle count and size distribution for large ferrous particles. The in-line design of the magnetometer enables both large and total ferrous content measurement. Eventually this eliminates the need another instrument for ferrous analysis.

    Oil Particle Counter Models Available:

    • SpectroLNF- Q210: LNF Oil Particle counter.
    • SpectroLNF -Q220: LNF Oil Particle counter with wear particle shape classifier.
    • SpectroLNF- Q230: LNF O,l Particle counter, automatic shape classifier, and ferrous monitoring with a magnetometer.
  2. Techical Information
    Analiz verileri Partikül sayım: ISO 4406, NAS 1638, NAVAIR 01-1A-17, SAE AS 4059, GOST, ASTM D6786, HAL ve kullanıcı tanımlı, serbest su, kurum wt. %; partikül şekli, partikül sınıflandırması, toplam ve büyük metalik partikül sayımı
    Standart ölçüm aralığı 4 μm – 100 μm
    Numune hacmi 5-30 mL
    Ortam gereksinimi 40°C ± 1°C
    Numune hacmi 2 mL
    Ortam gereksinimi 25°C – 35°C, %10 – %80 Bağıl nem
    Güç Gereksinimi 110/240 VAC, 50/60 Hz, 10 Watt
    Boyutlar (GxDxY) 43 x 18 x 23 cm
  3. Methods
    • ASTM D6786
    • ASTM D8120
    • ASTM D7596
    • ISO 4406
    • NAS 1638
    • NAVAIR 01-1A-17
    • SAE 4059
  4. Order Enquiry
  5. Documents
  6. Related Products