- General Details
Measuring CFPP Principle:
The sample is cooled down according to the methods and when the preselected temperature is reached a vacuum of 20 mBar is automatically applied to the sample. The product is sucked through the filter into the calibrated aspiration pipette. If the sample takes more than 60 seconds to reach the upper barrier detector (during the aspiration phase), or it fails to return completely into the test jar before that the product has cooled by a further 1°C, the Cold Filter Plugging Point is reached.
- Techical Info
Temperature range +80°C … -80°C Temperature stability 0.06 °C Precision ± 0.1 °C Cooling Single stage (for temperatures up to -40°C / 1)
Double stage (for temperatures up to -80°C / 2)
Requirements Temp: Max 32°C Power req. 220V ± 15% / 50 to 60 Hz 115V ± 15% / 60 Hz
Weight 70 kg Dimensions (WxDxH) 66 x 60 x 80 cm - Standards
- ASTM D6371
- IP 309
- IP 419
- EN 116
- EN 16329
- Order Enquiry