NL200 Cold Filter Plugging Point (CFPP) Analyzer

Automatic cold filter plugging point (CFPP) analysis of diesel, biodiesel and heating fuels.

Category:
  1. General Details

    Measuring CFPP Principle:

    The sample is cooled down according to the methods and when the preselected temperature is reached a vacuum of 20 mBar is automatically applied to the sample. The product is sucked through the filter into the calibrated aspiration pipette. If the sample takes more than 60 seconds to reach the upper barrier detector (during the aspiration phase), or it fails to return completely into the test jar before that the product has cooled by a further 1°C, the Cold Filter Plugging Point is reached.

  2. Techical Info
    Temperature range+80°C … -80°C
    Temperature stability0.06 °C
    Precision± 0.1 °C
    CoolingSingle stage (for temperatures up to -40°C / 1)
    Double stage (for temperatures up to -80°C / 2)
    RequirementsTemp: Max 32°C
    Power req.220V ± 15% / 50 to 60 Hz

    115V ± 15% / 60 Hz

    Weight70 kg
    Dimensions (WxDxH)66 x 60 x 80 cm
  3. Standards
    • ASTM D6371
    • IP 309
    • IP 419
    • EN 116
    • EN 16329
  4. Order Enquiry
  5. Related Products