- Home
- Products
- Petrochemistry
- OL580 Noack Tester
OL580 Noack Tester
- General Details
Procedure A uses the Noack evaporative tester equipment.
Procedure B uses the automated non-Woods metal Noack evaporative apparatus.Measuring Noack Principle:
A quantity of 65 grams of sample is heated to a specific temperature and maintained for 1 hour while it is enclosed in a crucible, the crucible’s cover is shaped to allow a constant vacuum of -2 mbar to remove from the crucible the evaporating portion of the sample.
At the end of the test, the sample is cooled and then reweighted: the difference, reported in percentage, represent the sample’s Evaporation Loss by the Noack Method.
Method A: bath is controlled at 250°C.
Method B: the sample is controlled at 245.2°C. - Techical Info
Test range +225°C to +275°C Measurement range 0°C … +320°C Temperature precision 0.01 °C Temperature stability ± 0.1 °C Heating Elektrikli, 500 W Power 220V ± 15% / 50 to 60 Hz
115V ± 15% / 60 HzWeight 22 kg Dimensions (WxDxH) 40 x 45 x 45 cm - Standards
- ASTM D5800
- IP 421
- Order Enquiry
- Documents