TAG Flash Point Tester

Flash point on petroleum products having a flash point between ambient temperature and +93°C.

Category:
  1. General Details

    Flash point on petroleum products having a flash point between ambient temperature and +93°C. Suitable for flash point detection on different substances and (NO) waste materials, solvents. The sample is warmed up according to the methods. When the sample reaches the selected test temperature, the shutter is opened and the ignition system introduces itself automatically. If the flash point is reached, the detection is done by an ionisation detector. If not, the shutter closes again and the sample continues to warm up until the next test temperature.


    Measuring Tag Devices

    • Measurement of the Flash Point detected by an ionisation detector
    • Testing unit equipped with two ignition systems
    • Electrical system or flame exposure device
    • Built-in barometric sensor with automatic barometric correction of results executed by the software
  2. Techical Info
    Measuring temperature probePlatinum resistance PT100 class A
    Measuring range-50°C … +100°C
    Testing range+9°C … +93°C as per test methods
    Resolution0.06 °C
    Accuracy± 0.1 °C
    Power supply220V ± 15% / 50 to 60 Hz
    115V ± 15% / 60 Hz
    Weight32 kg
    Dimensions (WxDxH)48 x 37 x 61 cm
  3. Standards
    • ASTM D56
    • ASTM D3278
    • ASTM D3934
    • ASTM D3941
    • IP 304
    • IP 491
    • IP 492
    • ISO 1516
    • ISO 1523
    • ISO 3679
  4. Order Enquiry
  5. Related Products