Minivap VP Vision Vapor Pressure Analyzer

Vapor pressure measurements of gasoline, jet fuel, crude oil, LPG and solvents with an extended pressure range of 0-2000 kPa.

Category:
  1. General Details

    MINIVAP VP Vision is a highly versatile portable vapor pressure tester that features best-in-class precision and a pressure range of 0-2000 kPa. It is the first analyzer that demonstrates excellence in engineering by earning certificates for robustness and durability. Based on Grabner´s cutting-edge Cockpit™ technology, the instrument offers unmatched networking capabilities and enables worldwide access to analyzers.

    MINIVAP VP Vision measures gasoline, jet fuel, crude oil, LPG and solvents covering an extended pressure range of 0-2000 kPa. The analyzer provides a wide pressure range, without compromising precision.

  2. Technical Info
    • Temperature Range:  Measured: 0 to 120°C (32 to 248°F), user programmable
    • Extrapolated: -100 to 300°C (-148 to 572°F)
    • Temperature Stability :+/- 0.01°C (0.018°F)
    • Temperature Profiles:  Single temperature, multipoint, curve, extrapolation
    • Pressure Range: 0 to 2000 kPa (0 to 290 psi)
    • Pressure Resolution:  0.01 kPa (0.0014 psi)
    • Precision Repeatability:  r ≤ 0.2 kPa (pure substance @ 37.8°C)
    • Reproducibility:  R ≤ 0.5 kPa (pure substance @ 37.8°C)
    • Vapor/Liquid Ratio:  0.02/1 to 100/1, depending on method
    • Measurement Time:  ~ 5 min., depending on method
    • Sample Introduction:  Automatic with built-in piston (no vacuum pump required)
    • Sample Volume:  1 mL (2.2 mL per rinsing cycle)
    • Results Database:  > 100,000 detailed results w/ graphics stored within the instrument
    • Results Download:  csv, PDF
    • Instrument OS:  Microsoft® Windows®
    • Instrument Electronics: Intel® Industry PC, 10” industrial touchscreen, 24 bit ADC
    • Instrument Languages:  en / de / fr / es / pt / it / ru / ar / zh / ja; customizeable
    • PC Software: Grabner Cockpit™ with automatic instrument recognition formulti-location results and user management, remote device
      configuration, update, diagnostics, support and calibration checks
    • Interfaces:  2x USB, 2x LAN (1x DVI-I, 1x RS 232) for direct connection to LIMS, PC, printer, LIMS, keyboard, mouse or barcode reader
    • Power Supply: 100-264 V AC, 45-63Hz,80W (Switching Power Supply)
    • Field Use: DC/AC Converter 12V (option)
    • Environmental, Shock, Vibration Certificates
    • EN 60068-2-1, EN 60068-2-78, EN 60068-2-14;
    • EN 60068-2-6, EN 60068-2-27 (IEC 60721-3-2, Class 2M2)
    • Dimensions:  (WxHxD) 293 x 390 x 280 mm (10.5 x 15.4 x 11 inch)
    • Weight: 10.5 kg (28 lb)
  3. Standards
    • ASTM D5191
    • ASTM D5188
    • ASTM D6377
    • ASTM D6378
    • ASTMD6897
    • EN 13016-1+2
    • IP 394, 409, 481
    • JIS K2258-2
    • SHT 0769, 0794
    • SNT 2932
    • GOST 52340
      • Excellent correlation to:
    • ASTM D323
    • ASTM D1267
    • ASTM D2879
    • ASTM D4953
    • ASTM D5482
      • Fuel Specifications
    • ASTM D910
    • ASTM 1655
    • ASTM 1835
    • ASTM 4814
    • ASTM 6227
    • EN 228
    • ASTM D5188 & ASTM D5191 Fast Mode™ Method Runs D5188 and D5191 in combination.
    • TVP Method
      Determines TVP at a V/L = 0/1.
    • Direct / Indirect VOC Methods For low volatile samples. Results correlate well to D2879 Isoteniscope method.
    • Customizeable Methods
  4. Order Inquiry